Infant home respiratory monitoring using pulse oximetry

Seddon, Paul, Sobowiec-Kouman, Sonia and Wertheim, David (2016) Infant home respiratory monitoring using pulse oximetry. Archives of Disease in Childhood, ISSN (print) 0003-9888 (Epub Ahead of Print)

Full text available as:
[img]
Preview
Text
Wertheim-D-35472-AAM.pdf - Accepted Version

Download (310kB) | Preview

Abstract

Respiratory rate (RR) is a valuable early marker of illness in vulnerable infants, but current monitoring methods are unsuitable for sustained home use. We have demonstrated accurate measurement of RR from brief recordings of pulse oximeter plethysmogram (pleth) trace in full-term neonates in hospital. This study assessed the feasibility of this method in preterm infants during overnight recordings in the home. We collected simultaneous overnight SpO2, pleth and respiratory inductive plethysmography (RIP) on 24 preterm infants in the home. RR from pleth analysis was compared with RR from RIP bands; pleth quality was assessed by the presence of visible artefact. Median (range) RR from RIP and pleth were not significantly different at 42 (25-65) and 42 (25-64) breaths/min. Median (range) % of epochs rejected due to artefact was 20 (8-75) for pleth and 10 (3-53) for RIP. Our results suggest that home RR monitoring by pulse oximeter pleth signal is accurate and feasible.

Item Type: Article
Additional Information: This work was supported by the National Institute for Health Research [grant number PB-PG-0610-22433].
Research Area: Allied health professions and studies
Computer science and informatics
Faculty, School or Research Centre: Faculty of Science, Engineering and Computing > School of Computing and Information Systems
Related URLs:
Depositing User: Automatic Import Agent
Date Deposited: 11 Jul 2016 08:11
Last Modified: 22 Jul 2016 08:04
URI: http://eprints.kingston.ac.uk/id/eprint/35472

Actions (Repository Editors)

Item Control Page Item Control Page